Objective To explore the comparability of alpha fetoprotein (AFP) and carcinoembryonic antigen (CEA) detection results and provide basis for result concordance in different chemiluminescence immunoassay systems. Methods Referring to the document of NCCLS EP9-A2, the testing Results derived from two analysis systems were analyzed comparatively and their biases was evaluated by analysis in SIEMENS ADVIA Centaur XP as a control method and analysis in BHP9504 as experimental method. Results The testing results of AFP and CEA by two different systems were linear correlation (r>0.99). Their anticipated biases were within accepted biases. Conclusion The test results of AFP and CEA by two different systems could be accepted by clinic.