Performance Verification of Sysmex UF-1000i Urinary Sediment Analyzer in the ISO15189 Accreditation

Expand
  • (Department of Clinical Laboratory,Cancer Hospital, Chinese Academy of Medical Science,Beijing 100021, China)

Received date: 2015-01-30

  Online published: 2015-09-16

Abstract

Abstract: Objective To evaluate the performance of Sysmex UF-1000i urine sediment analyzer in order to meet the requirement of the ISO15189 laboratory accreditation. Methods A series of tests with UF-1000i sedimentary analyzer of precision,accuracy,carryover and linear range were detected,as well as the biological reference interval was verified.Results Precision, accuracy, carryover and reportable range of UF-1000i were all in the allowable range supplied by factory authorities and the results of biological reference interval test met the requirements. Conclusion The performance of the Sysmex UF-1000i Urine Sedimentary analyzer was satisfied, it could be used for screening analysis of urine sedimentary.

Cite this article

ZHENG Cui-ling, QI Jun, WANG Li, XU Xiao-na, DONG Bo, CHENG Yan, HAN Xiao-hong . Performance Verification of Sysmex UF-1000i Urinary Sediment Analyzer in the ISO15189 Accreditation[J]. Labeled Immunoassays and Clinical Medicine, 2015 , 22(7) : 690 . DOI: 10.11748/bjmy.issn.1006-1703.2015.07.031