Comparison of Analysis Results Determined by Three Different Type Hematology Analyzers

Expand
  • (Clinical Laboratory, The First People's Hospital of Suzhou Wujiang Area,Suzhou 215200, China)

Received date: 2014-12-08

  Revised date: 2014-12-30

  Online published: 2015-09-16

Abstract

Abstract: Objective To compare the analysis results of three different type hematology analyzer to guarantee the consistency of laboratory data and promote the reliability of assay results. Methods The hematology analyzer Sysmex XS-1800i was set as reference. The WBC, RBC, HB, HCT and PLT from fresh blood samples with high, medium and low value daily were analyzed by Sysmex XS-1800i, Sysmex XN-2000 and Sysmex XS-800i analyzer. The linearity and recursion of data on the basis of 1/2 of external quality assessment permission error from American clinical laboratory amended draft (CLIA’88) were determined. Results The correlation index of assay data between different analyzer was more than 0.975 (P>0.05). The analysis results of three different hematology analyzers were correlated very well between each other. Conclusion The assay data of three different hematology analyzers are comparability and could guarantee the consistency of blood samples examination and reliability of assay test.

Cite this article

QIAN Jing . Comparison of Analysis Results Determined by Three Different Type Hematology Analyzers[J]. Labeled Immunoassays and Clinical Medicine, 2015 , 22(6) : 581 . DOI: 10.11748/bjmy.issn.1006-1703.2015.06.032