Comparison of β-HCG Detection Results by Two Analyzers

Expand
  • Department of Clinical Laboratory, The People’s Hospital of Pinghu, Shenzhen 518111, China

Online published: 2014-09-03

Abstract

Objective To explore the comparability of β-HCG detection results by two analyzers. Methods The serum β-HCG levels were detected by ABBOTT Architect i1000 and SIEMENS Dimension XPand analyzer. The results were compared by referring to the document of NCCLS EP9-A2. Results The testing results of serum β-HCG by two different analyzers were linear correlation (r>0.975).Their anticipated biases were within accepted biases (20%).Conclusion The two analyzers could be used for detection of serum β-HCG with clinical acceptance.

Cite this article

LUO Bao-min, LU Li-ping, HUANG Jin-yue, XIAO Yun-tong . Comparison of β-HCG Detection Results by Two Analyzers[J]. Labeled Immunoassays and Clinical Medicine, 2014 , 21(4) : 447 . DOI: 10.11748/bjmy.issn.1006-1703.2014.04.030