用极差检验方案验证两个检测系统天冬氨酸氨基转移酶测定结果的可比性
网络出版日期: 2013-12-14
Comparability of the Aspartate Aminotransferase Results Analyzed by Two Detection Systems and Verified through Range Test
Clinical Laboratory of
Online published: 2013-12-14
ZHAN Si-en , DENG Bao-min , SONG Yu-jie , ZHAI Yan-hong , WANG Wen-juan . 用极差检验方案验证两个检测系统天冬氨酸氨基转移酶测定结果的可比性[J]. 标记免疫分析与临床, 2013 , 20(6) : 440 -442 . DOI: 10.11748/bjmy.issn.1006-1703.2013.06.023
Abstract: Objective To explore the comparability of the AST results analyzed on two biochemistry analyzers and verified through range test. Methods According to the range test comparability scheme, two samples of different AST concentration were selected and analyzed respectively on two biochemistry analyzers to assess the comparability of the detection systems. Results Based on the evaluation standards of the comparable biological variability, the range value obtained by the two detection systems was under acceptable criteria. Conclusion The AST results analyzed by the two detection systems were absolutely comparable, and the range test comparability scheme is simple and eminently practical.
Key words:
Range test
/
| 〈 |
|
〉 |